The standard IEC 60364-7-710 requires an acoustic and visual alarm system to be installed at a suitable place so that the medical staff can permanently monitor it. In most cases, the remote signalling panel is installed on the operating theatre/room panel or at the nurse station.
Model | RMS-12 |
Displays | LCD |
Parameters Display | Insulation Resistance/Impedance, Load Current Transformer Winding Temperatures |
Alarm Signalling | Buzzer and LCD |
Buttons | 4 (esc (mute), up, down and enter) |
Serial Interface | RS 485 Modbus RTU |
Ethernet | Module RJ45 Ethernet 100 base-T |
Outputs | 2 Auxiliary Relays (NO-C-NC) |
Power Supply | AC 115...230 V or AC/DC 20...60 V or AC/DC 90...250 V, 50...60 Hz |